Notice: Function _load_textdomain_just_in_time was called incorrectly. Translation loading for the wordpress-seo domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /var/www/html/wp-includes/functions.php on line 6260 Cook Medical Recalls Hydrophilic Wire Guides for Raw Materials Contamination - Medical Product Outsourcing
Regulatory

Cook Medical Recalls Hydrophilic Wire Guides for Raw Materials Contamination

Adverse events due to contamination could include vessel damage, bleeding, and embolic particulate in the circulatory system.

DSM Biomedical B.V., the Netherlands, Cook Medical’s supplier of hydrophilic coating for the Roadrunner Uniglide Hydrophilic Wire Guide, recalled certain lots of material due to concerns about a potential contamination by glass particles ranging in size of approximately 4 to 280 µm. Cook received one lot of the affected recalled coating. This prompted Cook Medical to initiate a recall of 8,750 units of the Roadrunner UniGlide Hydrophilic Wire Guides on June 24, 2016. Cook is taking this action as a precautionary measure, because it cannot exclude the possibility that glass particles passed through during processing and could be present in the coating of the recalled wire guide units.
 
Potential adverse events that may occur as a result of this contamination could include vessel damage, bleeding and embolic particulate in the circulatory system. Cook Medical has not received any reports of adverse events of these types to date.
 
The Roadrunner UniGlide Hydrophilic Wire Guide is intended for use in the delivery of percutaneous catheters into the peripheral vasculature.
 
A full list of affected products and lot numbers is attached. Products can be identified by the part number and lot number on the outer label of the package. Products in this recall were distributed globally.
 
Cook Medical has notified its customers and distributors by recall notification letters. The letters requested that all customers and distributors quarantine and discontinue use of all recalled units and return the affected products to Cook as soon as possible for credit.
 
The FDA and other regulatory agencies around the world have been notified of this action.